Kaye IRTD 400 Intelligent RTD Secondary Standard

Calibration and Testing EquipmentThe intelligent resistant temperature device (IRTD) is a secondary standard with a completely self-contained measurement system, including electronics for calibration and temperature conversion. Kaye's IRTD provides ultra-high accuracy temperature data to your computer with no additional measurement or readout equipment required. The IRTD removes common sources of error because its integrated electronics and sensor are calibrated as a single instrument. Analog to digital conversion is automatic - no manual interpolations or calculations are necessary. The self-contained IRTD allows you to cut system error in half while eliminating the cost and effort of calibrating supporting equipment. For thermal validation applications, the IRTD communicates bath temperature directly with the Kaye systems.

The IRTD is an integral part of Kaye's automated sensor calibration system. In addition, the IRTD is invaluable for confirming the accuracy of other temperature standards. For this purpose, Kaye has developed IRTDWin™ software. This software automatically records data and computes any delta between two probes so you can easily confirm the accuracy of your traveling standard. Simply plug in your stationary and traveling IRTDs to your PC. The IRTDWin™ software displays and records the probes serial numbers and live temperature values. One convenient interface screen lets you track probe stability, numerically and graphically, and confirm system accuracy.

Tech Resources
Upgrade Download

Features

  • Temperature range -195 to 420°C
  • Accuracy over range ±0.025°C
  • Resolution 0.001°C
  • Sensor element 200 Ohm Platinum sensor
  • Sheath material Inconel™ 600
  • Immersion depth 101.6mm (4")
  • Calibration Traceable to NIST or PTB; Recommended calibration period is 1 year
  • Measurement rate 30 readings per second
  • Ambient temperature range 0 to 60°C (32 to 140°F)
  • Humidity Humidity 0 to 95% non-condensing
  • Dimensions Overall length: 603mm (23.75"); Grip: 89mm x 32mm (3.5" x 1.25") Sensor sheath: 457mm x 6.35mm (18" x 0.25")

Applications

Specifications

 