Calibration / Repair

Validation Calibration & Services

As the manufacturer we maintain laboratory accreditation (NVLAP according to ISO 17025 and DAkkS according to EN 17025) and provide both “mail-in” and “on-site” service options using our modern high precision calibration facilities and high end lab equipment.

We offer complete product support for the life of your Kaye equipment-going beyond typical calibration to provide customers with the highest quality of calibration and repair services. A timely calibration schedule assures accurate and reliable product performance and maximizes equipment uptime and productivity.

 

Americas

Kaye Validator- Factory Calibration

  • AS-FOUND, adjustment and AS-LEFT calibration with NIST traceable certification
  • SIMs: AS-FOUND, adjustment and ASLEFT calibration of the internal cold junction references (V2K calibration includes the calibration of up to 3 Sims)
  • Very low measurement uncertainties for calibration to guarantee high performance
  • Includes 8 hour elevated temperature soak to identify potential electronic latent defects
  • Free clock battery replacement included
  • Power loss test to check the back-up battery. If required a free exchange of backup battery
KG-052

Kaye IRTD-400 - Intelligent Temperature Standard - Factory Calibration

  • NVLAP AS-LEFT certificate includes -196°C, 0°C, 100°C and 400°C points
  • NVLAP AS-FOUND certificate includes -196°C, 0°C, 100°C, 165°C and 400°C points
  • Annealing Process at 400°C for a minimum of 15 hours to release sensor element stress and to offer more stability and improved performance between calibration cycles
  • We always adjust and reprogram the IRTD internal calibration coefficients over the full range
main_irtd400

Kaye Valprobe™ Wireless Process Validation and Monitoring - Factory Calibration

  • NVLAP AS-FOUND calibration at minimum 4 points for Temperature, Humidity & Pressure
  • High accuracy NVLAP AS-LEFT calibration at minimum 4 points
  • Adjustment of internal coefficients to achieve optimal reading
  • Includes free replacement of Lithium batteries and o-ring gaskets
ValProbeFamily

Kaye RF ValProbe™ Wireless Thermal Validation System - Factory Calibration

  • High accuracy NVLAP AS-LEFT calibration at minimum 4 points for Temperature and Humidity
  • NVLAP AS-FOUND calibration at minimum 4 points
  • Adjustment of internal coefficients to achieve optimal reading
  • Includes free replacement of Lithium batteries
main_RF_ValProbe

Kaye LTR/HTR Thermal References - Factory Calibration

  • AS-FOUND calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
  • Adjustment of internal calibration coefficients
  • AS-LEFT calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
  • Cycle and stability test at highest temperature, Cut-Out Safety Test
  • Free firmware upgrade upon request
main_htr_ltr_series

Europe

Kaye Validator - Factory Calibration

  • AS-FOUND, adjustment and AS-LEFT calibration with DAkkS traceable certification
  • SIMs: AS-FOUND, adjustment and ASLEFT calibration of the internal cold junction references (V2K calibration includes the calibration of up to 3 Sims)
  • Very low measurement uncertainties for calibration to guarantee high performance
  • Free clock battery replacement included
  • Verification of all 36 channels and communication ports
  • Internal clock verification
  • Power loss test to check the back-up battery. If required a free exchange of backup battery
KG-052

Kaye IRTD-400 - Intelligent Temperature Standard - Factory Calibration

  • DAkkS AS-LEFT certificate includes -196°C, -60°C, 0°C, 100°C, 260°C and 420°C points
  • DAkkS AS-FOUND certificate includes -196°C, -60°C, 0°C, 100°C, 260°C and 420°C points
  • Annealing Process at 420°C for a minimum of 70 hours to release sensor element stress and to offer more stability and improved performance between calibration cycles
  • We always adjust and reprogram the IRTD internal calibration coefficients over the full range to achieve optimal reading
main_irtd400

Kaye Valprobe™ Wireless Process Validation and Monitoring - Factory Calibration

  • DKD AS-FOUND calibration at minimum 4 points for Temperature, Humidity & Pressure
  • High accuracy DKD AS-LEFT calibration at minimum 4 points
  • Adjustment of internal coefficients to achieve optimal reading
  • Includes free replacement of Lithium batteries and o-ring gaskets
  • Logger maintenance, electrical tests
ValProbeFamily

Kaye RF ValProbe™ Wireless Thermal Validation System - Factory Calibration

  • High accuracy DKD AS-LEFT calibration at minimum 4 points for Temperature and Humidity
  • DKD AS-FOUND calibration at minimum 4 points
  • Adjustment of internal coefficients to achieve optimal reading
  • Includes free replacement of Lithium batteries
main_RF_ValProbe

Kaye LTR/HTR Thermal References - Factory Calibration

  • AS-FOUND calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
  • Adjustment of internal calibration coefficients
  • AS-LEFT calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
  • Cycle and stability test at highest temperature, Cut-Out Safety Test
  • Well to Well uniformity test at 121°C
main_htr_ltr_series
 